2-24
Close/Open Switching Module Channels
Model 2700 Multimeter/Switch System User’s Manual
Dual independent multiplexers
Using multiple channel operation, any multiplexer switching module can be configured as
two independent multiplexers. For example, the Model 7700 is normally used as a single
1
×
20 multiplexer, but it can also be configured as two 1
×
10 multiplexers.
NOTE
Thermocouple temperature measurements using the internal or external
reference junction cannot be performed when using multiple channel operation
to connect an input channel to the DMM. The simulated reference junction will
instead be used resulting in invalid readings (“ERR” annunciator turns on). See
“Temperature measurements,” page 3-33
, for details.
A multiplexer switching module is configured as two multiplexers by using multiple
channel operation to close the 2-pole/4-pole relay. The Model 7700 is configured as two
independent multiplexers by closing channel 23. As shown in
position of channel 23 isolates Multiplexer A (channels 1 through 10) from Multiplexer B
(channels 11 through 20).
For the dual multiplexer configuration, only Multiplexer A channels can be internally
connected to the DMM of the Model 2700. For the Model 7700, closing channel 25 allows
channels 1 through 10 to be measured by the DMM.
When using the dual multiplexer configuration, the sense backplane isolation relay must
be kept open to isolate Multiplexer B channels from the sense terminals of the DMM. For
the Model 7700, channel 24 must be kept open (
Figure 2-8
Dual multiplexer configuration (Model 7700)
Ch 1
Ch 10
Ch 11
Ch 20
Ch 23
(Closed)
Ch 24
Ch 25
Multiplexer A
(1x10)
Multiplexer B
(1x10)
For the dual multiplexer configuration,
Ch 23 must be closed, and Ch 24 must
remain open.
HI
LO
HI
LO
Channels
2–9
HI
LO
Channels
12–19
HI
LO
Sense
HI
LO
Input
To
Model 2700
DMM
HI
LO
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......