3-48
Basic DMM Operation
Model 2700 Multimeter/Switch System User’s Manual
Figure 3-19
Continuity connections
Continuity testing procedure
NOTE
Make sure the INPUTS switch is in the correct position. To use front panel
inputs, it must be in the “F” (out) position. For switching modules, it must be in
the “R” (in) position.
1.
Apply the resistance to be tested, and, if using a switching module, close the
appropriate channel.
2.
Press SHIFT and then CONT to display the present threshold LEVEL.
3.
Use
,
,
Δ
, and
∇
to key in the desired level (1 to 1000
Ω
), and press ENTER.
If the measured circuit is below the set threshold level, the instrument will beep and
display the resistance readings.
If the measured circuit is above the threshold level, the instrument will not beep
and either display the resistance reading or the message “OPEN”.
If the reading is below 1100
Ω
, it will be displayed. If the reading is 1100
Ω
or
above, “OPEN” will instead be displayed.
4.
To disable continuity testing, select a different function (i.e., press DCV).
NOTE
The beeper can be disabled using the SYSTem:BEEPer:STATe OFF command.
However, the beeper will automatically enable the next time the continuity
testing function is selected.
Limits and digital outputs cannot be used when testing continuity with the
continuity (CONT) function. If you need to use these operations, use the
Ω
2
function to test continuity.
H
L
CH 1-20
Model 7700
Switching
Module
Resistance
Under Test
Note: Source current flows from input high to input low.
Model 2700
Input HI
Input LO
Resistance
Under Test
A. Front panel connections
B. Model 7700 connections
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......