
7-18
Scanning
Model 2700 Multimeter/Switch System User’s Manual
Setting delay
and
, a delay (auto or manual) can be set by the user.
With auto delay selected, the delay period depends on function and range (
).
With manual delay selected, the delay period can be set from 0 secs to 99 hrs, 99 mins,
99.999 secs.
Perform the following steps to set auto or manual delay:
1.
With the instrument in the normal measurement display state, press SHIFT and
then DELAY.
2.
Press
Δ
or
∇
to display AUTO (auto delay) or MAN (manual delay) and press
ENTER.
3.
If you selected MAN, you will be prompted to set the delay in the hour/minute/
second time format. Use the
,
,
Δ
, and
∇
keys to set the delay.
NOTE
Pressing the AUTO key sets the manual delay to 0.001 sec.
4.
With the desired manual delay displayed, press ENTER.
For remote programming, the TRIGer:DELay <NRf> and TRIGger:DELay:AUTO <b>
commands are used to set the delay. See
for details.
NOTE
The delay for ratio and channel average can only be set using remote
programming (
).
Monitor channel
While in the normal measurement state, a scan list channel can be used to monitor
readings. When a channel is selected to be the monitor, it will assume the setup of the scan
list channel.
NOTE
If you change the setup while a monitor channel is closed, that setup will be
copied to that channel in the scan list.
When a scan is started, the first channel in the scan list will be briefly displayed. While the
scan is in progress, the display will only show the reading(s) for the monitor channel.
After the last channel is scanned, the scan will disable with the monitor channel closed.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......