Model 2700 Multimeter/Switch System User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-19
Remote programming — ratio and channel average
Ratio and channel average commands
The ratio and channel average are listed in
. Details on these commands follow
the table.
NOTE
Queries are not included in
. All the math commands are provided in
.
Enabling/disabling ratio or channel average
As with front panel operation, enabling ratio disables channel average and conversely,
enabling channel average disables ratio.
Ratio and channel average delay
:RATio:DELay or :CAVerage:DELay sets the delay between the two channel measure-
ments for the enabled calculation. This delay is applied after the trigger delay in the trigger
model (see
for details). This delay cannot be set from the front panel. The 0.5s
default delay keeps the relays from cycling too fast. Setting a shorter delay may shorten
the life of the relays.
It does not matter which of the two commands you use to set the delay. The set delay
affects both ratio and channel average.
Table 5-3
Ratio and channel average commands
Commands*
Description
Def
[SENSe[1]]
Optional root command.
:RATio[:STATe] <b> [, <clist>]
Enable/disable ratio; <b> = ON or OFF.
OFF
:RATio:DELay <NRf> [, <clist>]
Set delay (in secs); <NRf> = 0 to 99999.999.
0.5
:CAVerage[:STATe] <b> [, <clist>]
Enable/disable channel average; <b> = ON or OFF.
OFF
:CAVerage:DELay <NRf> [, <clist>]
Set delay (in secs); <NRf> = 0 to 99999.999.
0.5
Channel list parameter:
<clist> = (@SCH)
where: S =
Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
*The <clist> parameter is used to configure one or more channels for a scan.
Содержание 2700
Страница 3: ......
Страница 7: ......
Страница 197: ...5 24 Rel Math Ratio Channel Average dB Model 2700 Multimeter Switch System User s Manual...
Страница 213: ...6 16 Buffer Model 2700 Multimeter Switch System User s Manual...
Страница 289: ...9 18 Limits and Digital I O Model 2700 Multimeter Switch System User s Manual...
Страница 335: ...11 24 Status Structure Model 2700 Multimeter Switch System User s Manual...
Страница 336: ...12 CommonCommands...
Страница 344: ...13 SCPISignalOriented MeasurementCommands...
Страница 353: ...13 10 SCPI Signal Oriented Commands Model 2700 Multimeter Switch System User s Manual...
Страница 363: ...14 10 FORMat and Misc SYSTem Commands Model 2700 Multimeter Switch System User s Manual...
Страница 364: ...15 SCPIReferenceTables...
Страница 391: ...A 2 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 392: ...Model 2700 Multimeter Switch System User s Manual Specifications A 3...
Страница 393: ...A 4 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 394: ...Model 2700 Multimeter Switch System User s Manual Specifications A 5...
Страница 395: ...A 6 Specifications Model 2700 Multimeter Switch System User s Manual...
Страница 400: ...B Model7700ConnectionGuide...
Страница 411: ...B 12 Model 7700 Connection Guide Model 2700 Multimeter Switch System User s Manual...
Страница 412: ...C StatusandErrorMessages...
Страница 419: ...C 8 Status and Error Messages Model 2700 Multimeter Switch System User s Manual...
Страница 420: ...D SignalProcessing SequenceandDataFlow...
Страница 433: ...D 14 Signal Processing Sequence and Data Flow Model 2700 Multimeter Switch System User s Manual...
Страница 434: ...E MeasurementConsiderations...
Страница 443: ...E 10 Measurement Considerations Model 2700 Multimeter Switch System User s Manual...
Страница 454: ...G IEEE 488BusOverview...
Страница 469: ...G 16 IEEE 488 Bus Overview Model 2700 Multimeter Switch System User s Manual...
Страница 470: ...H KE2700InstrumentDriver Examples...
Страница 489: ...H 20 KE2700 Instrument Driver Examples Model 2700 Multimeter Switch System User s Manual...
Страница 499: ......