71
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
ƃƚƐƗƐƝƘƯ
Ɵƞ
ƭƛƕƚƢƠƞƜƐƓƝƘƢƝƞƙ
ơƞƒƜƕơƢƘƜƞơƢƘ
(
ƍżƁ
)
Information on
electromagnetic
compatibility (EMC)
Table 6
Recommended separation distances between portable and mobile RF communications equipment and
the TELE PACK X GI model TP 200
The TELE PACK X GI model TP 200 is intended for use in an electromagnetic environment in which the RF disturbances are controlled. The user of the device
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the device, according to the power output of the communications equipment as recommended below.
Rated maximum output power
of transmitter
[W]
Separation distance d
[
m
]
according to frequency of transmitter
150 kHz to < 80 MHz
D
80 MHz to < 800 MHz
D
800 MHz to 2.5 GHz
F
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters [m] can be estimated using the
equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts [W] according to the transmitter
manufacturer.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects and
people.