70
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
ƃƚƐƗƐƝƘƯ
Ɵƞ
ƭƛƕƚƢƠƞƜƐƓƝƘƢƝƞƙ
ơƞƒƜƕơƢƘƜƞơƢƘ
(
ƍżƁ
)
Information on
electromagnetic
compatibility (EMC)
Table 4
Guidance and manufacturer's declaration – electromagnetic immunity –
for equipment and systems that are not life-supporting
The TELE PACK X GI model TP 200 is designed for operation in the electromagnetic environment defined below.
The user of the device should make sure that it is used in such an environment.
Interference immunity
tests
EN/IEC 60601
test level
Conformity level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used no
closer to any part of the TELE PACK X GI model TP 200, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to < 80 MHz
3 V
rms
B
150 kHz to < 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
B
80 MHz to < 800 MHz
E
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the compliance level in
each frequency range
b
.
Interferences may occur in the vicinity of equipment marked with the
following symbol:
7
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the unit is used exceeds the applicable RF
compliance level above, the unit should be observed to verify normal operation. If unusual performance characteristics are observed, additional measures may
be necessary, such as a modified alignment or a different location for the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3
V/m.