Information on
electromagnetic
compatibility (EMC)
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The TELE PACK X GI model TP 200 is designed for operation in the electromagnetic environment defined below.
The user of the TELE PACK X GI model TP 200 should make sure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Conformity level
Electromagnetic environment – guidelines
Electrostatic discharge (ESD)
IEC 61000-4-2
± 4 kV contact
± 8 kV air
Complies
± 4 kV contact discharge
± 8 kV air discharge
Floors should be made of wood, concrete or ceramic tile.
If floors are covered with synthetic material, the relative
humidity must be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 2 kV for power supply
lines
± 1 kV for input/output lines
Mains power quality should be that of a typical commercial
or hospital environment.
Surges acc. to
IEC 61000-4-5
± 1 kV outer conductor – outer
conductor
± 2 kV outer conductor – earth
Complies
± 1 kV outer conductor –
outer conductor
± 2 kV outer conductor –
earth
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short-time
interruptions, and fluctuations
of the supply voltage acc. to
IEC 61000-4-11
<5 % U
T
*
(>95 % dip in U
T
)
for 1/2 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 seconds
Complies
<5 % U
T
(>95 % dip in U
T
)
for 1/2 cycle
Complies
40 % U
T
(60 % dip in U
T
)
for 5 cycles
Complies
70 % U
T
(30 % dip in U
T
)
for 25 cycles
Complies
<5 % U
T
(>95 % dip in U
T
)
for 5 seconds
Mains power quality should be that of a typical commercial
or hospital environment.
If the user of the device requests continued functioning
even when interruptions of the power supply occur, it
is recommended that the device be powered from an
uninterruptible power supply or a battery.
Magnetic field at the supply
frequency (50/60 Hz) acc. to
IEC 61000-4-8
3 A/m
Complies
3 A/m
The magnetic fields at the mains frequency should be at
the values typical of business or hospital environments.
* Note: U
T