4 DESCRIPTION OF CONTROLS
4-34
EM210F-1
l
Alarm screen
Fig. 4.37 Alarm screen
When any abnormality is detected, the background of the related item on this screen
turns to red.
AIR:
The compressed air pressure is insufficient.
RT:
The pressure in the compressed air reservoir is too low.
PIG:
The Pirani gauge is damaged.
RP1:
Rotary pump 1 is not normal.
RP2:
Rotary pump 2 is not normal.
DP-1 Water:
The cooling water temperature of oil diffusion pump 1 is too high.
DP-2 Heater:
The heater of oil diffusion pump 1 is damaged.
4.4.6g Valve Status screen
③
②
①
Fig. 4.38 Valve Status screen