4 DESCRIPTION OF CONTROLS
4-40
EM210F-1
4.4.6n Alignment Panel for Maintenance screen
⑥
①
⑦
⑤
④
②
③
Fig. 4.46 Alignment Panel for Maintenance screen
①
DEF Select
Gun:
Assigns the “Gun Shift” function to the SHIFT knobs (L1-
③
,
R1-
③
) and “Gun Tilt” function to the DEF/STIG knobs (L1-
⑩
,
R1-
⑤
).
Spot:
Assigns the spot alignment function to the DEF/STIG knobs
(L1-
⑩
, R1-
⑤
).
CLA:
Assigns the “Beam Shift” function to the SHIFT knobs (L1-
③
,
R1-
③
) and “Beam Tilt” function to the DEF/STIG knobs (L1-
⑩
, R1-
⑤
).
IS1:
Assigns the image shift 1function to the DEF/STIG knobs (L1-
⑩
, R1-
⑤
).
IS2:
Assigns the beam shift 2 function to the DEF/STIG knobs (L1-
⑩
, R1-
⑤
).
PLA:
Assigns the PL alignment function to the DEF/STIG knobs
(L1-
⑩
, R1-
⑤
).
CL
STIG:
Assigns the condenser lens stigmator function to the
DEF/STIG knobs (L1-
⑩
, R1-
⑤
).
OL STIG:
Assigns the objective lens stigmator function to the DEF/STIG
knobs (L1-
⑩
, R1-
⑤
).
IL STIG:
Assigns the intermediate lens stigmator function to the
DEF/STIG knobs (L1-
⑩
, R1-
⑤
).
DET:
Used for the detector of an optional attachment.
②
Compensator
Shift:
Assigns the “Shift Balance” function to the DEF/STIG knobs
(L1-
⑩
, R1-
⑤
).
Tilt:
Assigns the “Tilt Balance function to the DEF/STIG knobs
(L1-
⑩
, R1-
⑤
).
Angle:
Assigns the “Angle Balance” function to the DEF/STIG knobs
(L1-
⑩
, R1-
⑤
).
③
NTRL
Used to set the selected “Deflector” data to the neutral values.