5 OPERATION
EM210F-1
5-3
Emulsion
Film
Dispensing magazine
Receiving magazine
Cassette
Dispensing
magazine
Lid
Bottom
plate
Fig. 5.1 Loading films into the dispensing magazine Fig. 5.2 Magazines
5.2.3b Inserting/removing the magazines into/ from the camera chamber
1.
Make sure valve V1is closed.
2.
Open the nitrogen gas source valve and turn the camera chamber door
handle clockwise until it stops.
?
The door will open in 5 minutes.
3.
Open the camera chamber door and close the nitrogen source valve.
If the door does not open in five minutes after turning the door handle clockwise,
turn the handle fully counterclockwise, then turn the handle clockwise again and
wait for another five minutes. Repeat this turning and waiting until the door opens.
?
If the gas source valve is not open and/or no nitrogen gas is stored, the camera
chamber door does not open.
4.
Draw out the magazine stand by pulling the handle.
CAUTION
Do not touch the camera drive mechanism in the camera chamber.