5 OPERATION
EM210F-1
5-55
5.8.3c
EM-23080 evacuation system
Part name
Specification
Exchanging frequency
Gum hose
18
´
42
´
1000
Every 5 years
Regulating valve
INA-13-956-A-00
Every 5 years
Pressure meter
S-130
Every 5 years
Urethane tube
6.5
´
10, UF-25-10 blue
Every 5 years
Oil mist filter
DT-250-25A
Every 5 years
Gum hose
18
´
42
´
1000
Every 5 years
RP oil
MR-100-IL
Every 2 years
V-belt
Every 2 years
5.8.3d EM-23085 evacuation system
Part name
Specification
Exchanging frequency
TMP damper
PM006459-X
Every 3 years
Nylon tube
AS4-50-06 (4
´
6)
Every 5 years
Oil mist filter
DT-250-25A
Every 5 years
Gum hose
18
´
42
´
1000
Every 5 years
RP oil
MR-100-1L
Every2 years
V-belt
Every 2 years