4 DESCRIPTION OF CONTROLS
EM210F-1
4-29
l
Stage screen
This is used when a specimen holder is inserted in the microscope column.
①
③
②
Fig. 4.30 Stage screen
①
SHIFT
The
►
and
◄
buttons shift the specimen in the X directions and the
▲
and
▼
buttons
shift the specimen in the Y directions. The Z buttons shift the specimen in the Z
direction. The amount of the specimen shift per step can be varied in the STEP
boxes under the triangle buttons and the amount of the specimen tilt per step under
the Z buttons.
②
TILT
The X buttons tilt the specimen in the X directions and Y buttons in the Y directions.
The Y tilts is impossible when a single tilt specimen holder is used. The amount of
the tilt angle per step can be varied in the STEP box. The specimen can be rotated
using these buttons when a specimen rotation holder is used.
③
NTRL
Used to neutralize the values in the selected box (XY, Z, TILT X and TILT
Y/ROT).