SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
24
3
E03
Left Motor Fault
00
Left Motor Short Circuit
Check the left motor, connections and motor
cable.
01
Left Motor Open Circuit
02
Left Motor Connection Fault B-
03
Motor Terminal Connected to B+
04
Left Motor Voltage Fault
05
Left Motor Bridge Fault
06
Too Many Hardware Current
Limit Events
07
Current Offset Out of Range
08
Hardware Current Limit Fault
4
E04
Right Motor Fault
00
Right Motor Short Circuit
Check the right motor, connections and motor
cable.
01
Right Motor Open Circuit
02
Right Motor Connection Fault B-
03
Motor Terminal Connected to B+
04
Right Motor Voltage Fault
05
Right Motor Bridge Fault
06
Too Many Hardware Current
Limit Events
07
Current Offset Out of Range
08
Hardware Current Limit Fault
5
E05
Left Park Brake Fault
00
Left Park Brake Drive-Time Test
Failed
Check the left park brake connections and cable.
01
Left Park Brake Output Enabled
When Wheelchair Idle
02
Left Park Brake Output Did not
Enable When Entering Drive Mode
03
Left Park Brake fault during
power-up testing
04
Left park brake feedback low
during drive (park brake short)
NUMBER OF FLASHES
DIAGNOSTICS CODE
ERROR CODE DESCRIPTION
SUB CO
DE*
DET
A
ILS
OF
ERROR CODE
POSSIBLE SOLUTION