SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
25
6
E06
Right Park Brake Fault
00
Right Park Brake Drive-Time Test
Failed
Check the right park brake connections and
cable.
01
Right Park Brake Output Enabled
When Wheelchair Idle
02
Right Park Brake Output Did not
Enable When Entering Drive Mode
03
Right Park Brake fault during
power-up testing
04
Right park brake feedback low dur-
ing drive (park brake short)
7
E07
Remote Fault
00
Local SR Fault (CPU, EEPROM,
etc.)
Check the communications bus, connections and
wiring. Replace the remote.
01
Joystick fault at the remote
02
Speed pot fault at the remote
8
E08
Controller Fault
00
Controller fault
Check connections and wiring. Replace power
module.
01
RAM fault
02
ROM fault
03
CPU fault
04
EEPROM fault
05
Watchdog fault
06
Stack fault
07
Software fault
08
Power-up testing fault
09
Relay fault or precharge fault
10
Bridge fault or disable all fault
11
Electronics fault: Thermistor
12
Calibration setting fault
9
E09
Communications Fault
00
Remote connection lost
Check connections and wiring. Replace Bus
cable.
01
Low communication mode
10
E10
General Fault
00
General fault
Check all connections and wiring. Contact
Invacare Technical Service.
11
E11
Incompatible/incor-
rect Remote
00
Incompatible/incorrect Remote
Wrong type of remote connected. Ensure the
branding of the joystick matches that of
controller unit.
NUMBER OF FLASHES
DIAGNOSTICS CODE
ERROR CODE DESCRIPTION
SUB CO
DE*
DET
A
ILS
OF
ERROR CODE
POSSIBLE SOLUTION