Labfors 5 – Version for Microorganisms
Before Fermentation/Cultivation
22 January 2019
Page
145
of
205
2.
Insert the red banana plug into the connector on the sensor
head.
8.2.11 Connecting the pH Sensor
To connect the pH sensor, proceed as follows:
METTLER pH sensor
Procedure
1.
Remove the protective cap or the aluminium foil from the sen-
sor.
2.
Place the connector of the sensor cable onto the sensor head.
3.
Turn the ring on the connector and screw it tight.
Ensure the sensor cable is not twisted or buckled.
ATTENTION
The cable shield can be damaged by buckling or twisting. This
may lead to faulty measurements.