Labfors 5 – Version for Microorganisms
Technical Data
Page
200
of
205
22 January 2019
Variant High End
Gas
Gas flow control
Accuracy MFC
Air
1 MFC
± 0.3 %
(final value)
± 0.5 %
(measured value)
Gases
Gas flow control and gas
mix control
Air + O
2
2 MFC
Air + N
2
2 MFC
Air + O
2
+ N
2
3 MFC
Measuring ranges of MFCs
1
and rotameters
Vessel
max. WV
2
(Litres)
Measuring range Rotameter
L min
-1
(Variant Basic)
Measuring range MFC
L min
-1
(Variant Standard)
Measuring range MFC
L min
-1
(Variant High End)
1.2
0.25 up to 2.4
0.05 up to 2.5
0.025 up to 2.5
2.3
0.30 up to 4.7
0.10 up to 5.0
0.05 up to 5.0
5.0
0.50 up to 11.0
0.20 up to 10.0
0.10 up to 10.0
10.0
1.00 up to 22.0
0.40 up to 20.0
0.20 up to 20.0
1
)
Mass flow controllers
2
)
Working volume
13.6.6 Antifoam
Description
Value
Sensor (optional!)
Conductive with dosing needle
Adjustable in height
Control
Peristaltic pump AF”(Antifoam)
Range
0 / 100 % (ON/OFF)
Содержание Labfors 5