Labfors 5 – Version for Microorganisms
Cleaning and Maintenance
22 January 2019
Page
169
of
205
Procedure
1.
Insert an Allen key or a thin metal bar into the opening on the
upper end of the stirrer shaft.
This locks the stirrer in its position.
2.
Position an adjustable spanner on the two recesses on the up-
per end of the stirrer shaft.
3.
Unscrew the stirrer shaft with the spanner counter clockwise
from the drive shaft.
4.
Check the O-ring on the drive shaft for damages, replace as
necessary.
10.4 Cleaning and Storing Individual Parts
The procedure described here applies to the following individual
parts:
Vessel
Accessories such as blanking plugs, spargers, dip tubes, addi-
tion port adapters etc.
Reagent bottles