Labfors 5 – Version for Microorganisms
Technical Data
22 January 2019
Page
203
of
205
Pump hoses & flow rates
Description
Value
Material
PharMed BPT
Standard
iD: 1.0 mm
Wall thickness: 1.1 mm
Flow rate: 3.5 ml min
-1
Option 1
iD: 0.5 mm
Wall thickness: 1.15 mm
Flow rate: 1.2 ml min
-1
Option 2
iD: 2.5 mm
Wall thickness: 1,0 mm
Flow rate: 17.2 ml min
-1
Flow rates
All flow rates at 74 min
-1
(100 % rotation speed)
13.6.10 Sterile Filters
Description
Value
All filters
Autoclavable
Disposable
Inlet air filter
(red label)
1
Retention rate: 0.2 µm
Diameter: 37 mm
Inlet air filter
(no label)
2
Retention rate: 0.2 µm
Diameter: 25 mm
Exit gas filter
(green label)
3
Retention rate: 0.3 µm dry, 1.0 µm wet
Diameter: 37 mm
1
)
for vessels with DN 115 and 150
2
).
for vessels with DN 200
3
)
for all vessel sizes