Labfors 5 – Version for Microorganisms
Installation and Initial Operation
Page
102
of
205
22 January 2019
CAUTION
Risk of burns and loss of property due to elevated temperature!
The temperature control circuit will overheat without an inserted
temperature sensor and/or without liquid in the vessel. This can
lead to burns and loss of property.
14.
Couple the motor.
INFORMATION
The motor is controlled via the basic unit. Its cable is plugged in
during installation and will not be unplugged anymore during
routing operation of the equipment.
15.
Switch the equipment on at the main switch and wait until the
system is booted.
7.7.2 Filling the Water Circuit
To fill the water circuit, proceed as follows:
Procedure
1.
Set a low setpoint for parameter temperature (e.g. 10 °C), in
order to activate the water supply into the vessel jacket or cold
finger.
2.
Start the bioreactor.
All parameters except for
Temperature
remain switched off;
switch them off if necessary.
a) Temperature control system with electrical heater pad
and cold finger:
You should hear water flowing into the cold finger now.
b) Water temperature control system:
You should hear and see water flowing into the vessel jacket.
If using an exit gas cooler:
The water supply of the exit gas cooler is activated now.