Labfors 5 – Version for Microorganisms
Options
22 January 2019
Page
59
of
205
3.
Ensure the sensor is equipped with an intact O-ring, fit one if
necessary.
4.
Carefully screw the pressure sensor into the threaded
adapter.
5.
Screw the steel cap onto the female sensor connector.
After autoclaving
The exit gas filter (on the exit gas cooler) must be connected to the
pre-installed pressure hose of the solenoid valve. This ensures that
the exit gas is led via solenoid valve into the atmosphere i.e. into
the exit gas line, which has to be installed on-site by the provider.
The sensor cable must be connected to the pressure sensor, too.
Proceed as follows:
Procedure
1.
Fit the pressure hose from the solenoid valve to the exit gas
filter.
2.
Unscrew the steel cap from the cable plug on the sensor.
3.
Plug the male plug of the sensor cable into the female plug on
the sensor.
To do so, the red mark on each plug must be aligned.
Maintenance
Basically, the pressure sensor is maintenance-free. The recalibra-
tion interval depends on the operating conditions. However, annual
recalibration is recommended by the sensor manufacturer.
4.11 LabCIP, CIP/SIP Unit
The equipment is used for cleaning (CIP) and sterilisation (SIP) of
the bioreactor, provided there is a certain equipment configuration.