Labfors 5 – Version for Microorganisms
Cleaning and Maintenance
22 January 2019
Page
167
of
205
10.3.9 Removing the Pocket for Temperature Sensor (Pt100)
Proceed as follows:
Procedure
1.
Loosen both slotted screws next to the port.
Ensure the screws do not get lost.
2.
Push the pocket from the inside of the vessel top plate up, so
that it can be pulled out of the port from the outside of the top
plate.
Ensure the O-ring on the pocket does not get lost.
10.3.10 Removing the Sparger and the Dip Tube(s)
Straight spargers and dip tubes can be removed from the outside
of the vessel top plate. Curved spargers and dip tubes can only be
removed from the inside of the vessel top plate.
Since this equipment uses ring spargers and straight dip tubes,
removal from the inside of the vessel top plate is described here.
This means that the vessel top plate is already removed.
Proceed as follows:
Procedure
1.
Loosen the slotted screw on the clamping adapter.
2.
Carefully pull the sparger/dip tube from the bottom out of the
clamping adapter.
3.
Unscrew the clamping adapter out of the port by hand.
Ensure that O-ring does not get lost.
10.3.11 Removing the Baffles
The four baffles can be removed, once the vessel top plate is re-
moved.
Proceed as follows:
Procedure
1.
Loosen and remove the hexalobular screw from the vessel top
plate.
2.
Pull the baffle off and put aside.
Ensure that the O-ring does not get lost.
3.
Proceed the same way with the remaining baffles.