Labfors 5 – Version for Microorganisms
Options
Page
58
of
205
22 January 2019
If pressure control is switched off (parameter
Pressure
OFF
), the
valve automatically opens to prevent build-up of pressure in the
vessel.
For details about the touch screen software refer to the separate
operating manual.
The pressure sensor is mounted into a 19 mm port in the vessel
top plate. A threaded adapter is used for this (refer to chapter “Ac-
cessories” for details).
A plastic cap protects the steel membrane from damage. A
screwed on steel cap protects the female cable plug during auto-
claving.
1 Threaded steel cap
2 Hollow screw
3 Protective cap (plastic)
Before autoclaving
The pressure sensor is mounted into a 19 mm port in the vessel
top plate and autoclaved with the culture vessel.
Proceed as follows for the mounting:
Procedure
1.
Fit a threaded adapter into the 19 mm port in the vessel top
plate.
2.
Carefully remove the protective cap from the steel membrane
of the sensor.
ATTENTION
The steel membrane is very sensitive and can be damaged
when knocking it on solid objects or when rubbed.
Carefully mount the pressure sensor by hand. Do not use any
tool!
1
2
3