Labfors 5 – Version for Microorganisms
Before Fermentation/Cultivation
Page
120
of
205
22 January 2019
Mount the sensors in such a way that they cannot come in
contact with other components or the glass vessel.
If the mounting depth of is adjustable (mounting with electrode
holder/clamping adaptor), make sure the mounting depth is
set correctly prior to autoclaving, as later adjustment repre-
sents a contamination risk.
pH sensor
Calibrate the pH sensor on the operating panel before mounting
and autoclaving.
pO
2
sensor
Mount the pO
2
sensor in such a way that it has good access to the
flow and there is no risk of bubbles collecting. (Calibration is car-
ried out AFTER autoclaving!)
pH sensor & pO
2
sensor METTLER
Cover the sensor heads of METTLER pH sensor and pO2 sensors
with aluminium foil during autoclaving.
pH sensor & pO
2
sensor HAMILTON
ATTENTION
Risk of damage to the (HAMILTON) pH and pO
2
sensor. Cover-
ing the sensor heads with aluminium foil during autoclaving may
lead to water gathering under the film, thus damage the contacts
on the sensor head.
pH and pO
2
sensor heads should
NOT
be covered with alumin-
ium foil during autoclaving.
For details on the safety, technical data, usage and maintenance
requirements for the pH and pO
2
sensors, see the separate docu-
mentation provided by the manufacturers.
8.1.17.1 Calibrating the pH Sensor
For a reliable pH measurement, a 2-point calibration with an upper
and lower reference buffer must be carried out before each cultiva-
tion. The pH sensor must be calibrated before autoclaving. This is
carried out on the operating panel and described in detail in the
separate operating manual of the touch screen software.