Labfors 5 – Version for Microorganisms
Accessories
22 January 2019
Page
73
of
205
Addition port adapter Ø 12 mm
Inside-Ø
3.0 mm
Outside-Ø hose connection
5.0 mm
Muts be fitted with O-ring.
Thread is used for mounting in 12 mm / Pg13.5 port.
Addition port adapter Ø 12 mm
Inside-Ø
6.0 mm
Outside-Ø hose connection
8.5 mm
Must be fitted with O-ring before mounting.
Thread is used for mounting in 12 mm / Pg13.5 port.
Addition port adapter Ø 19 mm
Inside-Ø
3.0 mm
Outside-Ø hose connection
5.0 mm
Fitted with fixed O-ring.
Threaded adapter is used for mounting in 19 mm port.
5.11 Inoculation Needles
Inoculation needles are used for feeding liquids into the culture
vessel, which cannot be autoclaved with the culture vessel. These
liquids may be e.g. the inoculum or heat-sensitive reagents.
If using an inoculation needle, then a septum (inoculation mem-
brane) must be fitted along with a septum collar in the appropriate
port. The inoculation needle is connected with the reagent bottle
and autoclaved. The liquid, e.g. the inoculum, which shall to be
added into the culture vessel, is shortly before addition filled into
the reagent bottle under sterile condition. The septum is then