Labfors 5 – Version for Microorganisms
Technical Data
Page
198
of
205
22 January 2019
Impellers
Type / Number
Material
Rushton impellers with 6 blades
Standard: 2 pieces
Option: 3 pieces
316L stainless steel, electro pol-
ished
Vessel
A
B
C
2 L and 3.6 L / NW 115
46 mm
11 mm
11 mm
3.6 l and 7.5 L / NW 150
54 mm
11 mm
11 mm
13 L / NW 200
70 mm
13 mm
19 mm
13.6.4 Temperature
Description
Value
Sensor
Type: Pt100 1/3 DIN-B
Heating
Double walled vessels:
Water circulation in vessel jacket.
Pump and heating 500 W integrated in
basic unit.
Single walled vessels:
With electrical heater pad
Cooling
Double walled vessels:
With tap water via water circuit into vessel
jacket
Option: with circulation chiller
Single walled vessels:
With tap water via cold finger in vessel
Range of measure-
ment
-5 °C up to +150 °C
Содержание Labfors 5
Страница 208: ......