81
Configuring Module-Specific Settings
Configure Model U8979 Charge Unit Settings
This setting allows you to choose between voltage measurement and acceleration measurement
(charge-output or built-in pre-amplifier) for a channel.
[Voltage]
mode and
[Preamp]
mode use BNC connectors, whereas
[Charge]
mode uses miniature
connectors.
A channel can measure either one of them.
Model U8979 can automatically recognize TEDS-compliant* sensors.
*: Transducer electronic data sheet
WARNING
Setting the measurement mode to
[Preamp]
allows Model U8979 Charge Unit
to constantly provide power (3.5 mA, 22 V) to sensors. Set any measurement
mode other than
[Preamp]
or tum off the instrument before connecting a sensor
or probe with a BNC terminal to avoid an electric shock or damage to the
measurement target.
> [Channel] > [U8979]
to
3
6
1
2
1
Tap the
[Mode]
box, and then choose a measurement mode from the list.
Mode
Measurement target
Measurement sensitivity
Voltage
Voltage
–
Charge
Charge-output acceleration sensor
0.1 pC/(m/s
2
) to 10 pC/(m/s
2
)
Preamp
Acceleration sensor with a built-in pre-
amplifier
0.1 mV/(m/s
2
) to 10 mV/(m/s
2
)
2
(When setting mode to
[Preamp]
) Tap
[TEDS Load]
.
Acquires sensitivity of a connected sensor. However, the instrument can acquire sensitivity of TEDS-compliant
acceleration sensors with a built-in pre-amplifier only.
When sensor sensitivity has been acquired, it is automatically set.
Advanced Functions
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...