350
External Input and Output
3
Tap the
[OUT1]
and
[OUT2]
boxes in the
[External out]
area in turn, and then from the list,
choose a signal output action.
Choose a condition under which the instrument outputs a signal.
Judge(Pass)
Outputs a low-level signal when a pass judgment is given for the numerical
calculation result.
Judge(Fail)
Outputs a low-level signal when a fail judgment is given for the numerical calculation
result.
Error
Outputs a low-level signal when some kind of error or warning occurs.
Busy
Outputs a low-level signal while rejecting a START signal because the instrument is
in the busy state such as performing a measurement and saving data.
[Waiting for Trigger]
Outputs a low-level signal while waiting for a trigger.
The instrument continues outputting the signal for a pass/fail judgment (low-level output) until it starts the next
measurement.
The instrument can output various signals depending on its state.
Output signal
Open-drain output (with voltage output), active low
Output voltage
range
High level:
4.0 V to 5.0 V
Low level:
0 V to 0.5 V
Maximum input
voltage
50 V DC, 50 mA, 200 mW
GND
OUT1
OUT2
100
kΩ
10
kΩ
5 V
High
4.0 V to 5.0 V
Low
0 V to 0.5 V
Output period
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...