Examples of Configuring Waveform Calculation Settings
Example of an IIR low-pass filter (LPF) setting
This section explains how to configure the Z1 calculation setting to pass the CH1-1 data through a
2nd-order IIR low-pass filter with a cutoff frequency of 100 kHz (100000 Hz).
(With a sampling rate of 500 kHz, which means a sampling period of 2 µs)
> [Calculation] > [Waveform calculation]
2
3
1
4
1
Tap the
[Calculation]
button to set it to
[On]
.
2
Set
[Area]
to
[Whole]
.
3
Tap the channel to use for calculation,
[Z1]
, to set it to
[On]
.
4
Specify an arithmetic expression.
Tap the
[Formula]
box of
[Z1]
to display the arithmetic expression setting dialog box.