210
Examples of Configuring Waveform Calculation Settings
5
Enter an arithmetic expression.
SQR(MOV(CH(1,1)
∗
CH(1,1),200))
Number of sampling per period
(Specify 200 when measuring 50-Hz commercial power with a sampling rate of 10 kS/s)
6
5
6
Tap
[OK]
.
The arithmetic expression setting dialog box closes and the arithmetic expression you have entered appears.
7
Tap
[Execute]
to execute the calculation.
8
Press the
START
key to start a measurement.
The instrument have acquired a waveform, then displaying the calculated waveform.
HIOKI MR6000A966-03