122
Configuring the Pre-trigger and Post-trigger Settings
5.4 Configuring the Pre-trigger and Post-trigger
Settings
Pre-Trigger
You can record not only
waveforms that appear after a
start trigger has been activated,
but also those that appear before
the trigger is activated.
Setting of timing:
[Start]
,
[Start/Stop]
Also records data before the start trigger.
START
Specified pre-trigger
length
Recording length or
recording time
Post-Trigger
You can also record waveforms
that appear after the specified
recording length elapses from the
stop trigger point.
Setting of timing:
[Stop]
,
[Start/Stop]
Also records data after the stop trigger.
Stop
Specified post-trigger
length
Recording length or
recording time
The pre-trigger and post-trigger are expressed as a percentage of the recording length
(unit: percent).
•
Setting all the trigger sources (including analog and interval triggers) to
[Off]
disables pre-trigger
and post-trigger settings.
•
The setting of the recording length or the recording time is prioritized over the post-trigger setting.
Example
When the recording length is set at10,000 points and the post-trigger length at 30%, and
the stop trigger is activated after the instrument measures 8000 points of waveforms,
the instrument records only 2000 points of the waveforms after the stop trigger has been
activated.
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...