145
Triggering the Instrument Using CAN Signals (CAN Trigger)
(4) Tap the
[Comparison bit]
box, and then set the CAN-trigger pattern.
X
Ignores a signal.
0
The CAN-trigger condition for each CAN signal is satisfied when the signal is at a low
level.
1
The CAN-trigger condition for each CAN signal is satisfied when the signal is at a high
level.
The CAN-trigger conditions are defined, as with the logic-trigger conditions, depending on the
combination of the condition setting of satisfaction and the trigger pattern setting.
See “5.7 Triggering the Instrument Using Logic Signals (Logic Trigger)” (p. 138).
You can monitor CAN bus states by tapping
[START]
on the CAN trigger setting screen.
See “Monitor the state of the CAN/LIN bus.” (p. 289).
Configuring the
Trigger Settings
HIOKI MR6000A966-03