9
Setting Measurement Conditions
Using the envelope
> [Status] > [Condition]
1
2
3
1
Under
[Measurement method]
, tap
[Envelope]
to use the envelop.
Normal
Does not use the envelope.
Envelope
Uses the envelope.
• What is
Normal
?
The instrument records data at the specified sampling rate.
• What is
Envelope
?
At the specified recording interval, the instrument records the maximum and minimum values among data
points sampled within each specified recording interval by sampling an input signal at an over-sampling rate*
of 100 MS/s. Hence, even though a relatively longer recording interval is specified, peaks of fluctuations can
be recorded without being missed.
*:
Over-sampled data (indicated by blue dots
in the figure) are not saved.
• Values to be acquired using the envelope
Maximum
Minimum
Recording interval
Oversampling interval (10 ns)
: Over-sampled data
:
Stored data (the maximum and
minimum values)
: Waveform shown on the screen
A set of sampled data acquired using the envelope consists of two values: the maximum value and the
minimum value. These values are taken from the measured values acquired by sampling an input signal at an
over-sampling rate during the recording interval chosen in the
[Sampling]
box. The screen displays a vertically
wide waveform line. When saved in an external storage device, the data set that consists of the maximum and
minimum values is stored for a single instance of measurement.
Measurement Method
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...