Configuring Module-Specific Settings
Configuring Model MR8990 Digital Voltmeter Unit settings
> [Channel] > [MR8990]
to
1
6
1
Tap the area that includes
[Notch frequency]
.
The setting dialog box appears.
2
Tap the
[Notch frequency]
box, and then choose a power frequency from the list.
Choose a power frequency of your region.
50 Hz
Sets the period to 20 ms.
60 Hz
Sets the period to 16.67 ms.
An incorrect power frequency setting causes measured values to be unstable.
3
Tap the
[NLPC]
box, and then enter an integration time.
Define an integration time based on the power line cycle (PLC), which is the time equivalent to one period of
the power frequency.
0.1
to
0.9
,
1
to
10
,
20
,
30
,
40
,
50
,
60
,
70
,
80
,
90
,
100
Example:
When the power frequency is 50 Hz and NPLC is set at 10, the integration time is calculated to be
200 ms (= 20 ms × 10).
The instrument updates waveform data every 200 ms.
4
Tap the
[Response]
box, and then choose a data update interval from the list.
Off
Updates data at intervals of the integration time entered in the
[NPLC]
box.
On
Updates the data at high speed by calculating moving averages.
•
Updates data at intervals of 0.1 PLC when NPLC is set at 9 or less.
•
Updates data at intervals of 1 PLC when NPLC is set at 10 or more.
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...