Configuring the Trigger Timing Setting
5.3 Configuring the Trigger Timing Setting
Configure the waveform recording settings the instrument follows when it is triggered.
> [Trigger] > [Common]
1
Tap the
[Timing]
box, and then choose a trigger recording method from the list.
Start
Starts recording when the instrument is triggered, and stops the recording after the
instrument has acquired the recording length of waveforms.
Stop
Starts recording when you press the
START
key, and stops the recording when the
instrument is triggered.
Start/Stop
Starts recording when a start trigger is activated and records data until a stop trigger is
activated.
When Model U8975 4ch Analog Unit or MR8990 Digital Voltmeter Unit or both are included in the
measurement modules, the displayed trigger point may lag behind the actual trigger point by one
sample.