Triggering the Instrument at Regular Intervals (Interval Trigger)
5.8 Triggering the Instrument at Regular Intervals
(
Interval Trigger)
Start triggers can be activated at specified intervals. Setting the recording mode to
[Repeat]
allows
the instrument to record waveforms at regular intervals.
•
When using the pre-trigger, the instrument starts monitoring interval-trigger times after the first
pre-trigger time elapses since the start of measurement.
•
No start triggers are activated by any interval triggers while the instrument is filling the pre-trigger
memory. An interval trigger triggers the instrument while the instrument is waiting for a trigger
after the instrument has filled the pre-trigger memory.
•
Since the clock is internally corrected, displayed trigger times may not synchronize with the
intervals of the interval trigger.
> [Trigger] > [Common]
2
3
1
1
Tap
[Interval trigger]
.
2
Tap
[Interval trigger]
to set it to
[On]
.
3
Tap the
[h]
,
[min]
, and
[s]
boxes, and then enter a period of time for interval triggers.
An internal trigger is generated at the start of measurement, and triggers are repeatedly generated at the
specified intervals.
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...