170
Configuring the Numerical Calculation Settings
Setting
Description
Sample screen
Time to level
Pulse count
Target channel
(Analog, logic, real-time
waveform calculation)
Allows you to specify
calculation target channels.
Times
(Phase contrast only)
Detects the time when the
number of times a waveform
crossed the specified level
reaches the specified number.
Level
*
Detects the time when a
waveform crossed the level
specified here or a pulse count
that crossed the level. No
setting is available for logic
channels.
Slope
Detects a time when a
waveform crossed the level
specified here while rising or
a pulse count that crossed the
level while rising.
Detects the time when a
waveform crossed the level
specified here while falling or
a pulse count that crossed the
level while falling.
Filter(samples)
(
Off
,
10
to
10,000
)
Allows you to enter duration
used to determine that a
waveform crosses the specified
level.
Only after a waveform has
crossed the level and does not
cross the level again within
the specified filter duration,
the waveform is considered to
have crossed the level.
This is useful to eliminate false
level-crossing events caused
by noise.
Level at time
Target channel
(Analog, logic, real-time
waveform calculation)
Allows you to specify
calculation target channels.
Method
Allows you to set the time-
specifying method.
Time
Time
Allows you to enter a time for
obtaining a measured value
with the trigger point position
fixed at zero.
Calc. No. Calculation
No.
Allows you to choose a
calculation number when you
use numerical calculation
results.
You cannot specify any
number equal to or more than
the calculation number you
specified.
Specifying a range using
Segment 1 and Segment 2 is
not available.
*: You can set a level within the range of −9.9999E+29 to −1.0000E29, at zero, within the range of +1.0000E29 to
+9.9999E+29. A value to five significant figures can be specified (a value to 10 significant figures can be specified
for a time of the level at time setting).
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...