205
Operators of the Waveform Calculation and Calculation Results
b
i
:
i
th-point data in calculation results,
d
i
:
i
th-point data acquired across the source channel
Waveform calculation
type
Description
Exponential average
(AVEEXP)
This function requires an attenuation constant to be specified in the second parameter k.
Calculates an exponential average of repeatedly measured waveform data sets.
(First of repeated measurements)
A
d
1
,i
=
1
,i
A
1
,i
:
i
th-point data in exponential average results of the first
measurement among repeated measurements
d
1
,i
:
i
th-point data obtained across a channel of the first
measurement among repeated measurements
(Second or later of repeated measurements)
A
k
k
m i
m i
m i
,
,
,
d
A
/
=
+ −
−
(
(
)
)
1
1
m
: Number of repeated measurements
A
m,i
:
i
th-point data in exponential average results of the
m
th
measurement among repeated measurements
d
m,i
:
i
th-point data acquired across a channel of the
m
th
measurement among repeated measurements
k
:
Attenuation constant
Use an exponential average independently.
Concurrent use with other calculations causes an error.
W
aveform Calculation Function
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...