External Sampling (EXT.SMPL)
15.2 External Sampling (EXT.SMPL)
Externally inputting the signal allows you to set the sampling rate at any value.
How to an input signal
1
Connect the external sampling terminal of the instrument and the sampling signal-outputting
device with the SMB cable.
See “2.4 External Sampling (EXT.SMPL)” in the Quick Start Manual.
2
> [Status] > [Condition]
3
Tap the
[External sampling]
button to set it to
[On]
.
4
Tap the box to the right of the
[External sampling]
box, and then from the list, choose which
direction to used for reception of the external sampling signal.
To accept a rising edge:
[ ]
To accept a falling edge:
[ ]
To accept both a rising edge and falling edge:
[ & ]
5
Input pulse waves or square waves to the EXT.SMPL terminal. The waves have to be with a
high-level voltage of between 2.5 V and 10 V and a low-level voltage of between 0 V and 0.8 V.
Available voltage range
High level: 2.5 V to 10 V; low level: 0 V to 0.8 V
Pulse width
High and low level: 50 ns or more
Response frequency
10 MHz or less
Maximum input voltage
10 V DC
EXT. SMPL
600
Ω
10 pF
5 V
GND
3
kΩ
5 V
t
t
H
t
L
t
H
> 50 ns, t
L
> 50 ns, t > 100 ns
High
2.5 V to 10 V
Low
0 V to 0.8 V
•
If a sampling signal with 5 MHz or more is inputted, trigger points are delayed by one sample.
•
Using Model 8968 High Resolution Unit disables the anti-aliasing filter even when you set
[A.A.F.]
to
[On]
.
HIOKI MR6000A966-03
Содержание Memory HiCorder MR6000
Страница 2: ...HIOKI MR6000A966 03...
Страница 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Страница 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Страница 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Страница 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Страница 391: ...HIOKI MR6000A966 03...
Страница 392: ...HIOKI MR6000A966 03...