AD104-R2, AD104-R5
61 / 75
ba_aed104_8_e.doc
For details on how to set this mode, see command
COF
:
COF128
to
COF 140
continuous output foll. power On (not for bus operation ).
Example:
external triggering with automatical output
1. set parameter of the AED (
ASF,ICR
,...)
2. activate external Trigger with command
TRC1,1,0,20,5;
.
3. Define the ouputformat of the meas. values with command
i;
. "i" depending on binary output/
ASCII-output, see command COF, no parameter setting possible,i after each Trigger event the result is
given out automatically (without
MAV?;
command)
4. Stop the automatical measuring value output with
STP;
.command
5. nonvolatile storage of automatic output in the EEPROM with
TDD1;
command
6. starts again the automatical output with RES command. After each Trigger event the result is given out
automatically (without
MAV?;
command)
7. Stop the automatical output with
STP;
command. Parameter setting is possible
8. Deactivate automatical output with
COF3
; command
9. Nonvolatile storage of none automatic output (if wanted) with
TDD1;
10
.
Single measuring value output or single trigger output (if wanted) with
MSV?;
or
MAV?;