54 / 75
ba_aed104_8_e.doc
Command:
TDD(1);
Effect:
The settings you have changed in the working memory are stored in the EEPROM
nonvolatile.
ADR
Adress
ASF
Filter settings
BDR
Baud rate
COF
Configuration of the data output
CSM
Checksum of the data output
FMD
Filter mode
ICR
Measuring rate
STR
Terminator resistances on/off
TAS
Switch Gross/Net
TAV
Content of the tare memory
TEX
Output data delimiter
TRC
Trigger function
ZSE
Initial zero setting
ZTR
Automatic zero tracking
Command:
TDD(2);
Effect:
Transfer of the parameters from the EEPROM into the RAM. The parameters listed under
TDD1 are copied from the EEPROM into the RAM. This occurs automatically after reset and
power On.
Parameter protection without battery
Storing all settings such as filter cut-off frequency, measuring rate,
Baud rate, ...
RAM
ROM
EEPROM
TDD0
TDD1
TDD2
TDD0
ROM default values
User settings stored
protected against power failure
Current
user settings
Fig. 3.4-1:
Protecting the set parameters