AD104-R2, AD104-R5
37 / 75
ba_aed104_8_e.doc
The output format for the measured value must be set
previously
via the command
COF
.
The measured value is output related to the relevant measuring range. The measured value can be a gross or
net measured value (command TAS). This
COF
command generates answers of constant length.
The
output length
for the command
MSV?;
depends in this case upon the output format (see COF command):
Output format
AED answer
No. of characters
Binary 4 Byte
yyyy CRLF
(y – binär)
6
Binary 2 Byte
yy CRLF
(y – binär)
4
ASCII (
COF3;
)
xxxxxxxx CRLF
(x - ASCII)
10
ASCII (
COF9;
)
xxxxxxxx,xx,xxx CR LF (x - ASCII)
17
CR: Carriage Return, LF: Line Feed
-1001500,25,000,
Meas. value status optional
2 Byte adress (00...31) optional
8 Byte measuring value (pos. sign = Blank)
Delimiter can be set (command TEX)
Selected delimiter
Example:
Definition of a measuring value in ASCII format (
COF9
)