AD104-R2, AD104-R5
43 / 75
ba_aed104_8_e.doc
ICR
Internal Conversion Rate
(Output rate of measured values)
Range:
0...7
Factory setting:
0
Response time:
<10ms
Parameters:
1
Password protection:
none
Parameter protect.:
with command
TDD1
Input:
ICR(0...7);
Entry of the measuring rate as decimal number from 0...7
The integration time determines the output data rate of the measured values and thus also the response time to
the measured value query with the command
MSV?;
.
ICRx = Averaging over
2
x
measuring values ,
with x= 0...7
The following setting possibilities result from this:
ICR
Output rate Mv/s
0
100
1
50
2
25
3
12
4
6
5
3
6
2
7
1
Observe the baud rate setting when setting the measured value rate. A high baud rate must be set at high
measured value rates to avoid measured data losses (see command
COF
).
Query:
ICR?;
Effect:
Output of the set measuring rate (0...7)
Important Note
At ICR1 there is an especially good suppression of a 50 Hz mains frequency which may possibly cause
interference.