22 / 75
ba_aed104_8_e.doc
Bus mode: COF16 to COF 28
If the decimal number 16 is added to the above stated output formats COF0...COF12, then the AED is switched
into the bus output mode. A measured value is output. The AED switches over to the partially active mode (each
new measured value is stored in the output buffer but not output). The measured value is output on the bus with
the select command S..;.
Example:
2 AED’s in bus operation
Command
Effect
S98;
All AED are partially active (listening but not transmission)
COF18;
Output in 2 byte binary output
ICR0;
Maximum measuring rate
MSV?0;
Continuous measurement in AED
S01;
Read measured value from 1st AED, immediate
output
of current measured value
, without
measuring time
S02;
Read measured value from 2nd AED, when response from first AED has been received
completely. Immediate
output
of current measured value
, without measuring time
S01;
Read measured value from 1st AED, when response from second AED has been received
completely
S02;
Read measured value from 1st AED, when response from second AED has been received
completely
...
...
STP;
Stop measured value output
S01;
Poss. new setting for 1st AED
Binary measured value output without CRLF: COF32 to COF44
If the decimal number 32 is added to the above stated binary output formats
COF0...COF12
, the AED is
switched into the following output mode for the measured values.
In the
binary measured value output
, the end character CR LF is left out, so that only 2 or 4 characters per
measured value are output. This measure increases the output speed of the measured values.
Format
Length Sequence of the measured value output
COF32
4 Byte
MSB before LSB
LSB=0 (no status)
COF34
2 Byte
MSB before LSB
COF36
4 Byte
LSB before MSB
LSB=0 (no status)
COF38
2 Byte
LSB before MSB
COF40
4 Byte
MSB before LSB
LSB=status/CRC
COF44
4 Byte
LSB before MSB
LSB=status/CRC
2-wire bus mode: COF64 ... COF76 (do not use for variants AD104-R2 and AD104-R5)
If the decimal number 64 is added to the above stated output formats
COF0...COF12
, then the AED is switched
into the 2-wire bus mode. This means that the AED answers no longer with “0“ or “?“ on command inputs. The
answer with the parameter or in the case of
MSV?
with the measured value occurs only for command enquiries
(e.g.
ASF?
). The command
MSV?0;
(continuous measured value transmission) may no longer be used in this
case since otherwise it is no longer possible to stop this output (apart from supply voltage off).