AD104-R2, AD104-R5
11 / 75
ba_aed104_8_e.doc
The connection is made through a six-core cable. The maximum cable length is 15 m.
No bus mode
is possible
in this type of communications (no bus driver).
AD104 – R2
Y6 -Ub
Y5 -GND
Y12-TRG
Y10-RxD
Y8 - n.c.
Y9 -TxD
Y7 - n.c.
Fig.5.2:
Connections on the PCB (AD104-R2)
The connection scheme for the asynchronous interface results as follows:
AD104-R2
Master
Receiving line
RxD
TxD
transmission line
Transmission line
TxD
RxD
receiver line
Operating voltage
U
B
supply voltage (5.6V..15V
DC
)
Ground
GND
GND
Ground
Ext. trigger
TRG
external trigger signal
The levels on the RxD and TxD lines are RS-232 levels, whereby the quiescent-signal level is <-3V (Low).
External trigger signal:
Quiscent-signal level:
TRG= 0V (Low)
Active measuring:
TRG= Low-High edge (0V...5V)
If the external trigger is not used, the input remains open.
5.3
Serial Interface RS485 (Bus mode with AD104-R5)
Up to 32 AEDs can be connected to a common bus line through the RS-485 interface. The baud rate can be
selected between 1200 and 38400 baud in this version.
The following specification applies for the transmission of one character:
Start bit:
1
Number of data bits:
8
Parity bit:
none / even