50 / 75
ba_aed104_8_e.doc
ENU
Engineering Unit
(User engineering unit)
Range:
4 letters or numbers (ASCII characters)
Factory setting:
none
Response time:
Output:
<15ms
Input:
<40ms
Parameters:
1
Password protection:
none
Parameter protect.:
with input
Input:
ENU
("abcd");
Entry of a unit. An arbitrary unit with max. 4 characters can be entered. If less
than 4 characters are input, the entry is supplemented with blanks. The entered
unit is not appended to the measured value. The characters
must be entered in quotation marks (“...“).
Query:
ENU?;
Effect:
Output of the unit with 4 characters.