AD104-R2, AD104-R5
41 / 75
ba_aed104_8_e.doc
The FIR-Filter (FMD1) can be described with the following Table:
ASF
Filter
length
(Tabs)
Cut-off
frequency [Hz]
at -3dB
20dB
attenuation at
frequency in
Hz
40dB
attenuation at
frequency in
Hz
Attenuation in
the Stop band
[dB]
Stop band
[Hz]
1
12
7.6
17
23
50...100
>25
2
14
6.6
15
19
50...80
>20
3
16
6.2
14
17
50...90
>19
4
16
5.5
12.5
16
50...80
>17.5
5
18
4.7
11
14
45...80
>15
6
20
4
9.5
12
45...85
>12.5
7
22
3.5
8
10
40...85
>10
8
22
3
7
8
40...80
>8
The filter is switched off with
ASF0
.
The filters ASF6...ASF8 are limited additionally in bandwidth by averaging (ICR>4).
The transient recovery time of the filters is calculated from filter length (12...24) multiplied by the filter scan rate
of 10ms.
Settling time filter : 120ms ... 240ms
Mean value formation (ICR) does affect the total settling time of the measuring chain. The total settling time
depends additionally upon the mechanical construction of the transducer, the dead load of the scale and the
weight to be weighed.