AD104-R2, AD104-R5
45 / 75
ba_aed104_8_e.doc
TAV
Tare Value
(Set / read tare memory)
Range:
0...±8388607
Factory setting:
0
Response time:
<20ms
Parameters:
1
Password protection:
none
Parameter protect.:
with command
TDD1
Input:
TAV(±Tare value);
Enter tare value 7digit with sign (max. ±8 388 607).
This value is set off with the
LDW/LWT
characteristic
scaled with the parameter NOV (0...
NOV
). The tare
memory is cleared (contents = 0) after characteristic
entries with the commands
SZA, SFA
or
LDW, LWT
.
Query:
TAV?;
Effect:
The contents of the tare memory are output.
The tare value is converted to the NOV value.
Output format
measured value
at nominal load
Nominal
taring range at
NOV>0
Maximum
taring range at
NOV>0
Nominal
taring range at
NOV=0
Maximum
taring range at
NOV=0
2 Byte binary
±
NOV
value
±150%
NOV
value
±1000000
±8 388 607
4 Byte binary
±
NOV
value
±150%
NOV
value
±1000000
±8 388 607
ASCII
±
NOV
value
±150%
NOV
value
±1000000
±1 599 999
Example:
NOV3000
;
(Scaling the scale)
TAS1;
(Gross output switched on)
MSV?;
1500CRLF
(Measured value lies at 50% = nominal load of the scale)
TAR;
(Taring and switching over to net output)
TAV?;
1500CRLF
(Enquire tare value)
MSV?;
0CRLF
(Net measured valuet)
TAS?;
0CRLF
(Net is switched on)
TAS1;
0CRLF
(Switching over to gross)
MSV?;
3000CRLF
(Measured value is at 100% = nominal load of the scale)
TAV?;
1500CRLF
(Enquire tare value, unchanged)