20 / 75
ba_aed104_8_e.doc
COF
Configurate Output Format
(Ausgabeformat für die Messwertausgabe)
Range:
0...255
Factory setting:
9
Response time:
<15ms
Parameters:
1
Password protection:
none
Parameter protect.:
with command
TDD1
Input:
COF(0...255);
Input of the output format for measured value command
MSV
The possible formats and the decimal number to be entered for them are listed in the following Table. The
measured value output refers here to the set nominal value of the
AED
(see command
NOV
).
Output at max. capacity
NOV> 0
NOV= 0
2 Byte binary
NOV value
20000
4 Byte binary
NOV value
5120000
ASCII
NOV value
1000000
For the 2-bytes binary output, the NOV value must be < 30000, otherwise the measured value is output with
overflow or underflow (7fff
H
or 8000
H
). With NOV30000, the overload range is only still approx. 2700 digits.
Query:
COF?;
Effect:
Output of the selected output format as three-digit decimal number from 0...255