AD104-R2, AD104-R5
29 / 75
ba_aed104_8_e.doc
SZA
Sensor Zero Adjust
(Factory characteristic – zero point)
Range:
0...1,599999e6
Factory setting:
adjusted on unloaded transducer
Response time:
<15ms...4.2s
Parameters:
1
Password protection:
yes
Parameter protect.:
after input of
SFA
Input:
SZA;
(Response time: <4.2s)
Effect:
Using this command, the transducer electronic system measures an input signal
between ±2.5mV/V, stores the measured value as the zero point of the factory
characteristic, but only starts the calculation of the new characteristic after
entering the parameter for
SFA
.
Input:
SZA<zero value>;
(Response time: <15ms)
Effect:
Instead of causing the AED to measure the applied signal, the value is entered
here. The entered value is stored, but accounted for only after entry of the
parameter for
SFA
as the zero point of the factory characteristic.
Query:
SZA?;
(Response time <15ms)
Effect:
The value used in the AED for calculating the factory characteristic for the zero
point is output in ±7 digits (e.g. -0000345crlf).
Important Note
Entry the command SZA and SFA ore execute. The data will be calculated if both parameters were given in
pairs ore been measured.