Function:
The general function is described in Section 4.2.
The position of the trigger level P3 depends on the output scaling (
NOV
). At
NOV
=0 (scaling off) the trigger level
will be on the characteristic 0...1000000. At
NOV
>0 the trigger level will be in the range 0...
NOV
.
The response time P4 has to be adjusted, thus the settling time is not lower as of the used filter.
Example 1 :
ICR0, ASF3
Æ
total settling time >= 160 ms
Æ
P4 = 16 (= 16 * 10 ms, if ICR=0)
Example 2 :
ICR1, ASF3
Æ
total settling time >= 160 ms
Æ
P4 = 8 (= 8 * 20 ms, if ICR=1)
The measuring time P5 provide an additional averaging. (P5=6
Æ
averaging over 6 values).
The external trigger will only be released again when the output value has been formed (no re-trigger function).
Important Note
If the command COF (128..140) was used to select automatic output and the trigger function is activated, then
the AD104 will output the measured value once after triggering and subsequent measurement. Thus a
measured value query by means of the command MAV? will not be necessary. The connected computer only
needs to receive this measured value.